DPS50-M Series
Medical
60 Watts
Total Power:
30 - 60 Watts
Input Voltage:
127 - 300 VDC, 90 - 264 VAC
# of Outputs:
Single
Electrical Specs
Input
Input range
Frequency
Inrush current
Input current
Efficiency
EMI/RFI
Safety ground
leakage current
90 - 264 VAC (wide range),
127 - 300 VDC
47 - 440 Hz
<50A @ 230 VAC, cold start @ 25C
<2.0A rms @ 115 VAC; <1A @ 230 VAC
>80% @ full load typical, @115VAC
FCC Part 15, Subpart B Class B &
EN55022 (CISPR 22) Class B
275碌A(chǔ) max. @ 50/60 Hz, 264 VAC input
Output
Special Features
鈥?Medical safety approvals
鈥?CE Mark EMC & LVD
鈥?Universal AC input
鈥?Fully regulated output
鈥?Overload and short circuit protection
鈥?Internal remote sense
鈥?Constant voltage
鈥?High efficiency
鈥?High MTBF
鈥?IEC320 AC receptacle, 3 pin (type C14)
鈥?Built in EMI filter (CISPR 22 Class B)
鈥?LED power good indicator
鈥?Input power <74 watts
鈥?Complies with EN61000-3-2
Maximum Power (Po)
60W (DPS52-M, 30W)
Hold-up time
10 ms. typical at full load @ 115VAC
Overvoltage protection 20% - 45% above nominal output
maximum.
Latching type, recycle AC to reset.
Overload protection
Output short circuit protection
auto recover. Overload protected
@ 120 - 150% above maximum rating.
Cable/connector
DC cable with 2.5 mm center plug
barrel type
DC plug center +v
DC plug outer -v
Safety
UL
CSA
VDE
CE
UL60601-1
CSA-C22.2 NO. 601.1-M90
07501/EN60601-1 (IEC601)
EMC & LVD
Environmental
Operating temperature: 0擄C to +40擄C
ambient. Derate each output as 2.5% per
degree from 40擄 to 60擄C. -20擄C start up.
Storage temperature: -40擄C to +85擄C
Electromagnetic susceptability
designed to meet EN61000-4;
-2, -3, -4, -5, -6, -8, -11, level 3
Humidity: Operating; non-condensing; 10%
to 95% RH
Vibration: Three orthogonal axes, 3 sweeps
per axe at 1 oct/min. 2G peak 8 Hz to 500 Hz,
operational
MTBF demonstrated: >550,000 hours at full
load and 25擄C ambient conditions
190
Technical Support: (888) 41ASTEC or (407) 241-2752
Americas: (760) 930-4600
Europe (UK) 44 (1384) 842-211
Asia (HK) 852-2437-9662
rev 06.14.05