DPM 2
8.25mm (0.32鈥? Digit Height
Programmable Decimal Points
Auto-zero
Auto-polarity
200mV d.c. Full Scale Reading (F.S.R.)
3錕?frac12; Digit Subminiature LCD Module
The DPM 2 uses the latest miniaturisation techniques to produce a very compact meter. The snap-in integral bezel makes installation easy. For
single rail use, the DPM 2S features a built in negative rail generator, enabling the meter to measure a signal referenced to its own power supply
0V
.
SCALING
A potential divider may be used to alter the full scale
reading (F.S.R.) of the meter - see table.
NOTES
The meter will have to be
re-calibrated by adjusting
the calibration potentiometer
at the rear of the module.
+
Ra
INPUT
Rb
1M
Standard Meter
Single Rail Version
Accuracy (overall error) *
Linearity
Sample rate
Operating temperature range
Temperature stability DPM 2
DPM 2S
Supply voltage
DPM 2
DPM 2S
Supply current
DPM 2
DPM 2S
Input leakage current (Vin= 0V)
Stock Number
Min.
Typ.
0.1
3
200
100
9
5
150
250
1
DPM 2
DPM 2S
Specification
Max.
鹵1
50
14
7
10
3
10nF
IN HI
-
2
IN LO
0
7
3
% (鹵1 count)
count
samples/sec
擄C
ppm/擄C
V
碌
A
pA
Unit
Required F.S.R.
2V
20V
200V
2kV
200
碌
A
2mA
20mA
200mA
Note
910k
1M
1M
10M
0R
0R
0R
0R
Ra
100k
10k
1k
1k
1k
100R
10R
1R
Rb
* To ensure maximum accuracy, re-calibrate periodically.
CONNECTOR SOURCING GUIDE
METHOD
NOTE
Ensure that Ra is rated for high voltage use.
SUPPLIED WITH PRODUCT
DIMENSIONS
All dimensions in mm (inches)
35.0 (1.38)
22.3 (0.88)
17.0 (0.67)
9.70 (0.38)
a
b
c d
a.
b.
c.
d.
e.
f.
g.
h.
2.00
6.00
1.60
6.00
2.00
0.50
7.26
1.76
(0.08)
(0.23)
(0.06) max
(0.23)
(0.08)
(0.02)
(0.28)
(0.07)
Panel cut-out
33.0 x 15.0 (1.30 x 0.60)
Panel thickness
1.0 to 2.5 (0.04 to 0.1)
ON BOARD SOLDER LINKS
32.5 (1.27)
14.5 (0.57)
DP1
V+
GND
CAP+
CAP-
REF HI
REF LO
COM
IN HI
IN LO
V-
DP2
DP3
REF
10
1
h
g
10
1
e
f
PANEL FITTING
Locate the meter by passing it through the front of the panel cut-out and gently push until the rear of the bezel is flush with the panel (DO NOT PUSH ON
THE LCD). The snap-in lugs will now automatically hold the meter firmly in position.